足部角质层、上臂姿势及使用电解质湿巾对生物电阻抗法测量重度肥胖患者体成分的影响

The Impact of Feet Callosities, Arm Posture, and Usage of Electrolyte Wipes on Body Composition by Bioelectrical Impedance Analysis in Morbidly Obese Adults

作者

Jessica Roekenes, Magnus Strømmen, Bård Kulseng, et al.

Department of Public Health and General Practice, Faculty of Medicine, Norwegian University of Science and Technology, Trondheim, Norway, etc.

论文主要内容

论文结论
上肢测量姿势严重影响BIA测量体脂百分比的准确性,但足部角质层厚度及使用电解质湿巾对测量结果没有明显影响。进行InBody测试应注意上肢与躯干呈15°夹角,肥胖人群应避免上肢与躯干间皮肤的接触。
目的
评价足部角质层、上臂姿势、使用电解质湿巾对重度肥胖患者测量BIA体成分结果的影响。
对象
36例重度肥胖的患者,其中男性13例,年龄范围28-70岁,BMI 41.6 ± 4.3 kg/m2。
方法
使用人体成分分析仪InBody720测量体脂百分比。比较去除足部角质层前后,使用及不使用电解质湿巾,以及使用及不使用订制辅助衬板(衬板用于校正上臂姿势),几种情况下,体成分测量
结果